FDA rules

21 C.F.R. § 868.5095

Retrograde intubation device.

Current through eCFR issue date November 12, 2026.

Sections

(a) Identification. A retrograde intubation device is a prescription device used to perform retrograde intubation via the cricothyroid membrane. The device may contain or be labeled for use with guidewires and intubating catheters, in addition to needles (§ 868.5090), syringe (§ 880.5860 of this chapter), and hemostats (§ 878.4800 of this chapter).

(b) Classification. Class II (special controls). The special controls for this device are:

(1) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use, including the following:

(i) Wire guide tensile, flex, fracture, and corrosion testing;

(ii) Catheter tensile strength testing at likely points of failure;

(iii) Catheter kink radius testing;

(iv) Compatibility of device components that interact, including compatibility in connection, disconnection, and ability to transfer fluids;

(v) Dimensional validation;

(vi) Accuracy testing of markings; and

(vii) Validation of the maximum airway pressure.

(2) Performance data must support the shelf life of the device by demonstrating continued sterility, package integrity, and device functionality over the identified shelf life.

(3) The device must be demonstrated to be biocompatible.

(4) Labeling must include:

(i) Instructions for use; and

(ii) Package labels that clearly identify the minimum compatible size of endotracheal tube.

Source and amendment notes

Amendments

[86 FR 73678, Dec. 28, 2021]

This resource is educational and is not legal advice. Reading it does not create an attorney-client relationship. Laws and rules change, and the right answer depends on the facts of your unique situation.

Get Started

Ready to offload marketing compliance?

Send a message to our experienced legal team. We'll discuss your business objectives and develop a strategic plan tailored to your needs.

We manage a limited number of engagements so we consistently provide our clients with maximum value. Since your time is scarce, if we discover we are not the right firm for your situation, we will say so and point you toward someone who is.

Response within 24 hours
Berkeley Robinson - Founder