FDA rules

21 C.F.R. § 874.6010

External mechanical stimulator for the relief of congestion.

Current through eCFR issue date November 12, 2026.

Sections

(a) Identification. An external mechanical stimulator for the relief of congestion delivers vibrations to the sinus and nasal areas to relieve congestion.

(b) Classification. Class II (special controls). The special controls for this device are:

(1) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use, including verification of specified mechanical stimulation parameters.

(2) Performance data must demonstrate the electromagnetic compatibility, battery safety, and electrical safety of the device.

(3) Software verification, validation, and hazard analysis must be performed.

(4) The patient-contacting components of the device must be demonstrated to be biocompatible.

(5) Human factors testing must demonstrate that users can successfully use the device in the intended use environment based solely on its labeling and instructions for use.

(6) Labeling must include the following:

(i) Device specifications, including the frequency range, maximum output power, and power source; and

(ii) Explanation of the user interface.

Source and amendment notes

Amendments

[91 FR 70009, Nov. 6, 2026]

This resource is educational and is not legal advice. Reading it does not create an attorney-client relationship. Laws and rules change, and the right answer depends on the facts of your unique situation.

Get Started

Ready to offload marketing compliance?

Send a message to our experienced legal team. We'll discuss your business objectives and develop a strategic plan tailored to your needs.

We manage a limited number of engagements so we consistently provide our clients with maximum value. Since your time is scarce, if we discover we are not the right firm for your situation, we will say so and point you toward someone who is.

Response within 24 hours
Berkeley Robinson - Founder