FDA rules

21 C.F.R. § 880.6510

Whole room microbial reduction device.

Current through eCFR issue date November 12, 2026.

Sections

(a) Identification. A whole room microbial reduction device is a medical device to be used to reduce microbial load on medical device surfaces following cleaning and disinfection.

(b) Classification. Class II (special controls). The special controls for this device are:

(1) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:

(i) Performance testing must demonstrate microbial log reduction of the demonstrated most resistant microorganism on medical device surfaces commensurate with the intended level of microbial reduction;

(ii) Simulated use testing must evaluate device performance under simulated worst-case use conditions (e.g., soiling, room objects and surfaces, distances);

(iii) In-use testing must evaluate device performance under real-world use conditions;

(iv) Performance testing must demonstrate the photobiological safety of any lamps or lamp systems;

(v) Performance testing must evaluate safety features intended to prevent exposure and ensure that device operation can only occur in an unoccupied environment; and

(vi) Performance testing must characterize the long-term material compatibility of the microbiocidal agent on clinically relevant surfaces and/or devices.

(2) Biocompatibility testing must demonstrate safe residual levels of chemicals on medical devices surfaces and/or gaseous byproducts in air.

(3) Software verification, validation, and hazard analysis must be performed for any software components.

(4) Performance data must demonstrate the electromagnetic compatibility and electrical safety of the device.

(5) Labeling must include:

(i) Warnings and instructions to ensure the device is operated in an unoccupied environment;

(ii) Setup and positioning instructions; and

(iii) Information regarding material compatibility.

Source and amendment notes

Amendments

[91 FR 57508, Sept. 10, 2026]

This resource is educational and is not legal advice. Reading it does not create an attorney-client relationship. Laws and rules change, and the right answer depends on the facts of your unique situation.

Get Started

Ready to offload marketing compliance?

Send a message to our experienced legal team. We'll discuss your business objectives and develop a strategic plan tailored to your needs.

We manage a limited number of engagements so we consistently provide our clients with maximum value. Since your time is scarce, if we discover we are not the right firm for your situation, we will say so and point you toward someone who is.

Response within 24 hours
Berkeley Robinson - Founder